HYDRAULIC INSTRUMENTS
H14/2 Computer Compatible Manometer Bank - Issue 2
The Armfield H14/2 is designed to replace banks of manometers when used in conjunction with a number of Armfield products. Sixteen simultaneous pressure measurements can be displayed on a user supplied computer and the information data logged.
Two modes of operation are available:
Mode 1: 16 general purpose channels (350mm H20), with a common reference, suitable for monitoring both wet and dry systems.
Mode 2: 14 general purpose channels as above,
plus 2 differential high sensitivity channels (125mm H20 differential),
for accurate measurement of low pressures (suitable for dry air only).
ORDERING SPECIFICATION
• Electronic multiple channel pressure measurement system designed
to replace liquid manometer banks.
• 16 simultaneous measurements displayed on a PC (not supplied).
• Sophisticated software included, for logging, recording and displaying
data.
• 16 channels of 350mm H20 capacity, or 14 channels of 350mm plus 2 channels
of 125mm H20.
Description
The eighteen pressure monitor points are located on the front panel of
the unit, with adapters to allow flexible plastic tubing to be attached.
Two different sizes of tubing can be used (2mm and 3mm inside diameter).
This tubing connects to the test points on the equipment being monitored.
The manometer system connects to the customer supplied computer using a
USB interface. The measured pressures are displayed on the computer screen
in both numeric and bar chart format.
Sophisticated software is provided with facilities to:
> Display the data from the 16 channels in real time
> Record data either on operator request, or at constant sampling periods
> Define the data sampling periods to be used
> Record data onto the computer for lateral retrieval, including the option
to record the data in Excel format
> Display tables and graphs of the captured data, with real time updates
> Calibrate and scale the individual measurement channels
> Zero the data channels before taking measurements
Specification
General purpose channels:
Range: 0 to 350mm H20
Repeatability: 0.5mm H20
Absolute accuracy: 0.5% FSD
Media: Wet or dry, non-corrosive
High sensitivity channels:
Range: 0 to 125mm H20
Repeatability: 0.1mm H20
Absolute accuracy: <0.5% FSD
Media: Dry air only
(all figures typical)
Essential Equipment
PC with USB interface, running Windows 98 or later.
Associated Armfield Products
C2: Subsonic Wind Tunnel
F6: Air Flow Studies
requirements
AC mains: Universal input, 100-250V ac, 50-60Hz
Shipping specification
No Data Available
Overall dimensions
Length: 0.25m
Width: 0.1m
Depth: 0.04m
|
![]() |
|---|
DATA SHEET
PRODUCT CATEGORY:
Engineering Education
H Series
Vernier hook & point gauges
- H1-1/H1-2/H1-3
Electronic hook & point gauges
- H1-7/H1-8
Liquid Manometers & Meters
- H12-1/H12-2/H12-3
H12-4/H12-5
Electronic pressure meter
- H12-8
Electronic pressure meter
- H12-9
Computer compatible manometer bank
- H14/2
Propeller velocity flowmeter
- H33